SPC Benzene Data Exercise
Ron Graham
with charliew@hal-pc.org
The data are from a benzene analyzer, measuring percentage
volume of benzene in gasoline. The measurement that is
displayed in the plot is [(analyzer feedback) - (a known
standard sample value)], which centers the numbers about
zero. Due to the very low variability shown (relative to
the repeatability of the analyzer), there is some
"jumpiness" to the data.
There are quite a few points in the first 3/4 of the
period that were above zero, indicating that we had a
"run rule" violation (more than seven consecutive
points above or below the line). At the end of the
data run, this situation was corrected, with the
instrument calibration adjusted. Here is the analysis
log:
- 01/31/2001 --
Being worked by Casillas, Stewart, and Adams. Instrument
has been calibration checked a couple of times and is still
above the mean. South and Offline charts are both OK.
- 02/07/2001 --
Checked calibration again - still OK. Changed out
calibration standard - also OK.
- 02/08/2001 --
Re-ran calibration on both North and South - still OK.
- 02/11/2001 --
Continue to watch. The South chart and our GC SQC chart
both look good. Casillas, Stewart, and Adams are looking
into the problem.
- 02/15/2001 --
Note downward trend, but still above the mean. Will
continue to watch.
- 02/20/2001 --
Finally dipped below the mean.